■ Characteristic ● Water-jacket heating and breeze circulation ensure temperature more uniform. Classical mode of heat transfer and uniform temperature. ● Easy operation: running with data set and time set, and automatic shutdown. ● Special function key for temperature setting. ● Auxiliary menu makes over-temperature alarm, deviation calibration, and menu buttons lock come true. ● Vertical double air flue air-casing, forced air convection, better temperature conformity. ● Forced air intake structure for fan motor cooling ensure its max working temperature less than 50℃ which lengthen the service life of fan motor. ● Warning water level alarm keeps the experiment safe. ● Double doors design reduce heat loss ● Inner reinforced glass door is convenient to observe the experiment status at any time. ● Patent No.: ZL201320059701.3 |
Technical parameters | ||||
Model | GH3000 | GH4500 | GH6000 | |
Application | bacteria incubator with water jacket and fan combination design | |||
Air circulation type | water-jacket heating, double air flues, and forced convection | |||
Performance | temp. range | RT+5~65℃ | ||
temp. resolution | 0.1℃ | |||
temp. fluctuation | ±0.5℃ | |||
temp. uniformity | ±1.0℃ | |||
Structure | chamber material | stainless steel | ||
shell material | spraying cold rolled steel | |||
thermal insulation material | PU | |||
heating device | stainless steel heating tube | |||
rated power | 0.45kw | 0.65kw | 0.85kw | |
vent | no | |||
Controller | temp. control | double rows of digital tube, PID | ||
temp. setting | press the four keys slightly | |||
temp. display |
actual temp. is displayed by 3 digit digital tube (row 1); setting temp. is displayed by 3 digit digital tube (row 2). |
|||
timer | 0~9999 minutes (with timing function) | |||
operation function | running with data set and time set, and automatic shut-down based on setting | |||
program mode | optional | |||
additional function | deviation calibration, menu buttons lock, power outage compensation, power outage memory. | |||
sensor | Pt100 | |||
Safety device | over-temperature alarm, warning water level alarm | |||
Specification | working size (W*D*Hmm) | 300*300*450 | 450*450*550 | 600*600*700 |
exterior size (W*D*Hmm) | 420*420*690 | 570*570*840 | 720*720*990 | |
packing size (W*D*Hmm) | 590*550*810 | 730*690*950 | 880*840*1100 | |
working capacity | 41L | 111L | 252L | |
shelf load bearing | 15kg | |||
max shelf layer | 8 layers | 11layers | 15 layers | |
height between shelves | 40mm | |||
Power source (50/60Hz) /Rated current | AC220V/2.0A | AC220V/3.0A | AC220V/3.9A | |
NW/GW | 30/42kg | 45/62kg | 70/88kg | |
Accessory | shelf | 2pcs | ||
shelf bracket | 4pcs | |||
Optional device | shelf, RS485 port, printer, recorder, external communication, remote control, program temperature controller, wireless message alert, U disk storage. |
上一篇:Electrothermal Constant Temperature Incubator (DH)